Ivoclar Programat P95 Manual

Users can select between standard Ivoclar firing programs or create individual custom programs. Calibration: Periodic temperature calibration (often via the Silver Test

If restorations come out over-fired (too glossy) or under-fired (matte, chipping), run a calibration test. The manual outlines using the "Calib" program with a ceramic calibration pellet. The target temperature difference should be less than 5°C. If higher, adjust the "Temp. Offset" parameter in the service menu (accessible with code 1234). Ivoclar Programat P95 Manual

A unique feature of the P95 is its powerful vacuum pump. The manual dedicates several pages to: Users can select between standard Ivoclar firing programs

The manual details the classic method of temperature calibration using silver wire , a process where the user must visually decide if the furnace is too hot or cold based on when the wire melts. The target temperature difference should be less than 5°C